Giếng chuẩn nhiệt PYROS 375
Portable temperature calibrators PYROS 375
– Applications:
+ Control and calibration of temperature sensor,
+ in the laboratory and in the field,
+ in conformity with ISO 9000 standards Control of thermostats.
– Automatic computer-controlled calibrations
Technical specification
– Operating range (t amb. 20 °C/68 °F): amb +10 °C ÷ +375 °C (Amb. +18 °F ÷ 707 °F)
– Mean heating time (stabilization included): From 30 °C / 86 °F to 375 °C / 707 °F 20 min
– Mean cooling time (stabilization included): From 375 °C / 707 °F to 100 °C / 212 °F 40 min
– Axial uniformity: ±0,10 °C at 50 °C
: ±0,20 °C at 150 °C
: ±0,30 °C at 375 °C
– Radial uniformity (at 40 mm)
: ±0,10 °C at 50 °C
: ±0,15 °C at 150 °C
: ±0,20 °C at 375 °C
– Hole depth: 150 mm (5.9 in.)
– Hole diameter: 26 mm (1 in.)
– Display accuracy (±1 digit): ±0,25 °C at 150 °C
– Stability: ±0,15 °C – (±0.27 °F)
– Display Resolution: 0,1 °C – (0.1 °F)
– Units of measure: °C / °F
– Over temperature cut out: Yes
– Switch test: 5 VAC DC
– PC Interface: RS 232
– Display: LED double line
– IN / OUT: RS232
– Temperature ramps: Yes
– Calibration: From 1 to 10 calibration points
– Calibrator weight: 5,4 kg – (11 lb)
– Power supply: Automatic switchable voltage supply 115/230 VAC ±10% – 50/60 Hz
– Electric power: 600 W
– Rigid case
+ Size: 410 x 535 x 220 mm (16 x 21 x 8.6 in.)
+ Weight: 3 kg (6 lb)
+ Shipping size: 470 x 640 x 330 mm
– Shoulder bag
+ Size: 400 x 170 x 285 mm (15 x 6.6 x 11 in.)
+ Weight: 1 kg (2 lb)
+ Shipping size: 450 x 250 x 415 mm (17 x 9 x 16 in.)
Standard equipment:
– Electric power cable.
– Tweezers for insert removing.
– Kit of fuses.
– Thermostat connection cables.